Are you struggling when having to tranfer a validated HPLC method between labs or different instruments made by different vendors?Transferring a validated method from one HPLC system to another can be a very demanding, complicated, time and money consuming procedure without guaranteeing optimum results.
When transferring methods between HPLC instruments made by different vendors, you need to consider that every system on the market has its own unique parameters, such as gradient delay and extra column volumes, plus column and solvent temperature control.
Thermo Scientific Vanquish HPLC and UHPLC Systems contributes to achieving the transfer smoother by providing the appropriate tools:
- Tunable GDV from 0 – 430 µL. A tunable GDV that physically mimics the delay volume of your original system to avoid retention time shifts, altered peak shapes, and selectivity. Click here and scroll down at the end of the page for an interactive GDV adjusting demo.
- Dual thermostatting modes to maintain analyte selectivity.
- Custom injection programs to sustain peak shape & gradient profile.
- Strong Solvent loop in order to preserve peak quality
- Third-party CDS integrations available with the Chromeleon Chromatography Data Systems.
Learn more about the unique method transfer capabillities of Thermo Scientific Vanquish HPLC and UHPLC Systems here